Original Article

Comparison of ocular biometry in Chinese patients using two swept-source optical coherence tomography devices

:22-77
 
Background: To compare two swept-source optical coherence tomography (SS-OCT) biometers,IOLMaster 700 and ANTERION.
Methods: This is a retrospective study. Biometric measurements of cataract patients performed between March and July 2021 in the Department of Ophthalmology, United Christian Hospital, Hong Kong, were reviewed. Patients scheduled for cataract surgery were measured with both SS-OCT devices on the same day.The following biometry parameters were compared: keratometry (K), total keratometry (TK), axial length (AL), central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), white-to-white (WTW) and the predicted intraocular lens (IOL) power to achieve emmetropia. To assess the agreement between the devices, Bland-Altman analysis with 95% limits of agreement (LoA) were used.
Results: In total, 92 eyes of 47 subjects were measured with both devices. There were statistically significant differences between the two biometers for most measurements (P<0.05) except for flat K, AL and IOL power when using the right eyes for analysis. For the left eyes, there were statistically significant differences in the measurements from the two biometers in all parameters except for flat and steep K. The ANTERION did not obtain ACD, AL and LT in 2 (2.17%), 1 (1.09%) and 5 cases (5.43%) respectively.
Conclusions: The two biometers showed a clinically acceptable agreement in most parameters. Comparisons showed significant differences in most parameters but not clinically relevant except for the TK and WTW, and these two parameters should not be used interchangeably between the devices.
Original Article

Changes in crystalline lens parameters during accommodation evaluated using swept source anterior segment optical coherence tomography

:-
 

Backgrounds: To assess changes in anterior segment biometry during accommodation using a swept source anterior segment optical coherence tomography (SS-OCT).

Methods: One hundred-forty participants were consecutively recruited in the current study. Each participant underwent SS-OCT scanning at 0 and ?3 diopter (D) accommodative stress after refractive compensation, and ocular parameters including anterior chamber depth (ACD), anterior and posterior lens curvature, lens thickness (LT) and lens diameter were recorded. Anterior segment length (ASL) was defined as ACD plus LT. Lens central point (LCP) was defined as ACD plus half of the LT. The accommodative response was calculated as changes in total optical power during accommodation.

Results: Compared to non-accommodative status, ACD (2.952±0.402 vs. 2.904±0.382 mm, P<0.001), anterior (10.771±1.801 vs. 10.086±1.571 mm, P<0.001) and posterior lens curvature (5.894±0.435 vs. 5.767±0.420 mm, P<0.001), lens diameter (9.829±0.338 vs. 9.695±0.358 mm, P<0.001) and LCP (4.925±0.274 vs. 4.900±0.259 mm, P=0.010) tended to decreased and LT thickened (9.829±0.338 vs. 9.695±0.358 mm, P<0.001), while ASL (6.903±0.279 vs. 6.898±0.268 mm, P=0.568) did not change significantly during accommodation. Younger age (β=0.029, 95% CI: 0.020 to 0.038, P<0.001) and larger anterior lens curvature (β=?0.071, 95% CI: ?0.138 to ?0.003, P=0.040) were associated with accommodation induced greater steeping amplitude of anterior lens curvature. The optical eye power at 0 and ?3 D accommodative stress was 62.486±2.284 and 63.274±2.290 D, respectively (P<0.001). Age was an independent factor of accommodative response (β=?0.027, 95% CI: ?0.038 to ?0.016, P<0.001).

Conclusions: During ?3 D accommodative stress, the anterior and posterior lens curvature steepened, followed by thickened LT, fronted LCP and shallowed ACD. The accommodative response of ?3 D stimulus is age-dependent.

Original Article

Subjective refractions determined by Dyop? and LogMAR chart as fixation targets

:-
 

Background: Dyop® is a dynamic optotype with a rotating and segmented visual stimulus. It can be used for visual acuity and refractive error measurement. The objective of the study was to compare refractive error measurement using the Dyop® acuity and LogMAR E charts.

Methods: Fifty subjects aged 18 or above with aided visual acuity better than 6/12 were recruited. Refractive error was measured by subjective refraction methods using the Dyop® acuity chart and LogMAR E charts and the duration of measurement compared. Thibo’s notation was used to represent the refractive error obtained for analysis.

Results: There was no significant difference in terms of spherical equivalent (M) (P=0.96) or J0 (P=0.78) and J45 (P=0.51) components measured using the Dyop® acuity and LogMAR E charts. However, subjective refraction measurement was significantly faster using the Dyop® acuity chart (t=4.46, P<0.05), with an average measurement time of 419.90±91.17 versus 452.04±74.71 seconds using the LogMAR E chart.

Conclusions: Accuracy of refractive error measurement using a Dyop® chart was comparable with use of a LogMAR E chart. The dynamic optotype Dyop® could be considered as an alternative fixation target to be used in subjective refraction.

Review Article

Psychophysics in the ophthalmological practice—I. visual acuity

:-
 
Perception is the ability to see, hear, or become aware of external stimuli through the senses. Visual stimuli are electromagnetic waves that interact with the eye and elicit a sensation. Sensations, indeed, imply the detection, resolution, and recognition of objects and images, and their accuracy depends on the integrity of the visual system. In clinical practice, evaluating the integrity of the visual system relies greatly on the assessment of visual acuity, that is to say on the capacity to identify a signal. Visual acuity, indeed, is of utmost importance for diagnosing and monitoring ophthalmological diseases. Visual acuity is a function that detects the presence of a stimulation (a signal) and resolves its detail(s). This is the case of a symbol like “E”: the stimulus is detected, then it is resolved as three horizontal bars and a vertical bar. In fact, within the clinical setting visual acuity is usually measured with alphanumeric symbols and is a three-step process that involves not only detection and resolution, but, due to the semantic content of letters and numbers, their recognition. Along with subjective (psychophysical) procedures, objective methods that do not require the active participation of the observer have been proposed to estimate visual acuity in non-collaborating subjects, malingerers, or toddlers. This paper aims to explain the psychophysical rationale underlying the measurement of visual acuity and revise the most common procedures used for its assessment.
Editorial
Editorial
Original Article
Theme 4: OMICS

AB015. The relationship between copy number variations and high myopia in Chinese: a case-control study

:-
 

Abstract: As a complex disease, myopia is the most common eye disease worldwide. Many myopia susceptibility genes or variants have been successfully identified in the past years by genome-wide genetic association studies (GWAS), which focus mainly on the single-nucleotide polymorphisms. Little attention has been paid to examine the role of copy number variations (CNVs) in refractive error and myopia. This study adopted a systematic strategy to investigate the role of CNVs in high myopia. In the discovery phase, a pilot GWAS suggests putative CNVs for follow-up. Multiplex ligation-dependent probe amplification was then used to quantify the copy number of 89 CNV segments in 737 case-control samples in the second phase and then 24 top-ranking CNVs in a second group of 1,029 case-control samples in the final validation phase. This validation phase identified 22 significant CNVs. Further work is needed to examine the role of these few CNVs in myopia development.

Theme 2: Ocular Development

AB008: Structural and molecular changes in cornea and sclera of highly myopic-astigmatic chicks

:-
 

Abstract: Myopia and astigmatism, two common refractive errors frequently co-exist, are degrading vision at all working distances in populations worldwide. Eyeballs having high degrees of myopia and astigmatism are known to exhibit abnormal eye shape at the anterior and posterior eye segments, but whether the outer coats of these abnormal eyeballs, cornea anteriorly and sclera posteriorly, are regulated by region-specific molecular mechanism remains unclear. Here we presented the changes in eye shape and mRNA expression levels of three genes (MMP2, TIMP2, and TGFB2), all known to participate in extracellular matrix organization, at five regions of the cornea and sclera in chickens developing high myopia and astigmatism induced by form deprivation. Our results showed that, compared to normal chicks, the highly myopic-astigmatic chicks had significantly astigmatic cornea, deeper anterior chamber, longer axial length, and higher expressions of all three genes in the superior sclera. These results imply that local molecular mechanism may manipulate the eye’s structural remodeling across the globe during refractive eye growth.

Theme 2: Ocular Development

AB007. Visual signals modulate refractive error development through dopamine receptor signaling

:-
 

Abstract: Myopia prevalence is dramatically increasing in recent years and in cases in which the refractive error is greater than ?6.00 D this disease can lead to severe visual impairment as well as even blindness. Changes in visual input affect the balance between ocular growth and refractive power development. If a mismatch occurs during eye development, the severity of this error affects the degree of myopia. In different animal models of this disease, we found that spatial visual stimuli are essential for maintaining a stable refractive status and normal vision. This is evident because the effects of changes in temporal visual stimuli (e.g., flickering light) on this process depend on whether spatial information is present or absent in the visual environment. Furthermore, the frequency, wavelength and intensity of light are involved in controlling refraction development. However, the molecular mechanisms underlying light-induced refraction changes are still unclear. There is definitive evidence that dopamine (DA) is one of the regulators of this process. This retinal neurotransmitter released by dopaminergic amacrine cells appears to play an important role in vision-guided eye growth because its synthesis and release are positively associated with the light intensity and spatial stimuli impinging on the retina. We found that bright light enhances retinal DA synthesis, and attenuates form deprivation myopia (FDM) development via activation of the dopamine receptor 1 (D1R). A nonselective DA receptor agonist apomorphine (APO) inhibited FDM in dopamine receptor 2 (D2R) knockout mice. These individual similar effects of DA and APO in wildtype and D2R knockout mice suggest that D1R activation has a protective effect against myopia development. On the other hand, D2R activation instead appears to promote myopia development because either genetic D2R ablation or pharmacological inactivation of D2R also attenuates myopia development. Based on these results, we hypothesize that the visual environment regulates the retinal DA levels, which in turn affects the relative balance between D1R and D2R activation. When D1R is relatively hyperactivated, the ocular refractive status shifts towards hyperopia. In contrast, such an effect on D2Rpromotes the refractive status to shift in the opposite direction towards myopia.

其他期刊
  • 眼科学报

    主管:中华人民共和国教育部
    主办:中山大学
    承办:中山大学中山眼科中心
    主编:林浩添
    主管:中华人民共和国教育部
    主办:中山大学
    浏览
  • Eye Science

    主管:中华人民共和国教育部
    主办:中山大学
    承办:中山大学中山眼科中心
    主编:林浩添
    主管:中华人民共和国教育部
    主办:中山大学
    浏览
出版者信息