泪膜改变对 Goldmann 眼压测量值的影响

Effect of Tear Film Changes on the Intr aocular Pr essur e Measur ement by Goldmann Applantion Tonometer

:27-61
 
目的: 探讨泪膜改变对 Goldmann 压平眼压计测量值的影响。
方法: 将 68 例(136 只眼) 受检者随机分为 2 组, 每组 34 例。随机选择一只眼作为实验眼, 另一只眼作为对照眼。A 组的实验眼采用右旋糖苷 70 滴眼液滴眼, B 组的实验眼采用粘弹剂 Viscoat 滴眼。采用 Goldmann 眼压计测量 2 次眼压, 比较滴眼前、后测量值的差异。
结果: A 组对照眼第 1、2 次的眼压测量值分别为(14.44±2.68) mmHg(1 mmHg = 0.133 KPa) 、(14.47 ± 2.69) mmHg, 两次眼压测量值的差异无统计学意义(= - 0.329, = 0.744) 。实验眼滴右旋糖苷 70 滴眼液前、后的眼压测量值分别为(14.41 ±2 .63) mmHg、(12.94 ± 2.59) mmHg, 两次眼压测量值的差异有统计学意义 (t = 13.949, P = 0.000) 。B 组对照眼第 1、2 次的眼压测量值分别为(14.29 ± 2.96) mmHg、(14.35 ± 3.12) mmHg, 两次眼压测量值的差异无统计学意义(t = - 0.466, P = 0.644) 。实验眼滴 Viscoat 前、后的眼压测量值分别为 (14.53 ± 3.13) mmHg、(11.18 ± 3.07) mmHg, 两次眼压测量值的差异有统计学意义( t = 22.126, P = 0.000) 。两组的实验眼滴眼后的眼压均呈一致性的下降。
结论: 泪膜的改变可以使 Goldmann 压平眼压计的测量值产生偏差。
Purpose: To explore the effect of tear film changes on the intraocular pressure (IOP) measurement by Goldmann applantion tonometer.
Methods: Sixty-eight normal subjects were randomly divided into two groups (Gruop A and B) , 34 in each group. One eye of each subject by random selection was given Dextran (in Group A) and Viscoat (in Group B) , respectively. And the fellow eyes were used as controls. IOP was measured in all subjects twice using Goldmann applantion tonometer. And the difference between first and second measurements was compared.
Result: In the control eyes of Group A, the average IOP of first and second measurements were (14.44 ± 2.68) mmHg (1 mmHg= 0.133 KPa) and (14.47 ± 2.69) mmHg, there was no difference between the two measurements (t =- 0.329, = 0.744) . In eyes given Dextran, the average IOP of two measurements were (14.41 ± 2.63) mmHg and (12.94 ± 2.59) mmHg, there was significant difference between the two mea-surements (t =13.949, P= 0.000) . In control eyes of Group B, the average IOP of first and second measurements were (14.53 ± 3.13) mmHg and (11.18 ± 3.07) mmHg, and there was no difference between the two measurements (t = -0.466, = 0.644) . In eyes given Viscoat, the average IOP of two measurements were (14.53 ± 3.13) mmHg and (11.18 ± 3.07) mmHg, and there was significant difference between the two measurements (= 22.126, = 0.000) . After being given Dextran or Viscoat, the IOP values decreased consistently in both Group A and Group B.
Conclusions: The change of tear film components can affect IOP values by Goldmann applantion tonometer.