Objective: To compare the consistency of Pentacamand iTrace in detecting anglekappa. Methods: A retrospective study was carried out on a total of 172 eyes of 86 volunteers to measure the anglekappa by iTrace visual analyzer and Pentacamanterior section analyzer in Zhengzhou Aier Eye Hospital. The measured data were collected and analyzed by SPSS software. Results: In the measurement results between Pentacam and iTrace , there was a statistically significant difference in the X value, Y value, and azimuth angle values of the horizontal axis (P<0.05); There was no statistically significant difference in chord distance values (P>0.05). Conclusions: 3-D anterior section analyzer the significant difference is found in the concordance of the anglekappa measurement results between the Pentacam and iTrace. It should be caution to replace two kinds of equipment in clinical practice.